MODBUS Register Map v1.48, Rx104P
248083336-1 Rev 3 A-15
These registers run in parallel with the data registers (30xxx series).
For example, data register 30010’s enable register would be 43010.
Data register 30016’s enable register would be 43016.
Note:
Alarm Enable
currently only works for
Particle Channels.
The user can enable multiple particle channels for alarming at the same
time.
Particle data registers for the Enable setting start at 43009 for the high
word and 43010 for the low word for particle channel 1.
Table A-14 Alarm Enable Registers
Register Data Type Description
43009 unsigned int Enable for Particle Channel 1 [high] (smallest
particle size starts here)
43010 unsigned int Enable for Particle Channel 1 [low]
43011 unsigned int Enable for Particle Channel 2 [high]
43012 unsigned int Enable for Particle Channel 2 [low]
43013 unsigned int Enable for Particle Channel 3 [high]
43014 unsigned int Enable for Particle Channel 3 [low]
43015 unsigned int Enable for Particle Channel 4 [high]
43016 unsigned int Enable for Particle Channel 4 [low]
43017 unsigned int Enable for Particle Channel 5 [high]
43018 unsigned int Enable for Particle Channel 5 [low]
43019 unsigned int Enable for Particle Channel 6 [high]
43020 unsigned int Enable for Particle Channel 6 [low]
43021 unsigned int Enable for Particle Channel 7 [high]
43022 unsigned int Enable for Particle Channel 7 [low]
43023 unsigned int Enable for Particle Channel 8 [high]
43024 unsigned int Enable for Particle Channel 8 [low]
43041 unsigned int Enable for Analog Channel 1 [high]
43042 unsigned int Enable for Analog Channel 1 [low]
43043 unsigned int Enable for Analog Channel 2 [high]
43044 unsigned int Enable for Analog Channel 2 [low]