Methods
RESULT
Chlorine dioxide with Tablet
0.02 – 11 mg/l ClO
2
Select the method Chlorine CL 6 for determining the concentra-
tion of chlorine dioxide.
a) in absence of Chlorine
Fill a clean vial (24 mm Ø) with 10 ml of the water sample and
perform zero calibration (see “Operation”).
Remove the vial from the sample chamber and empty it, leaving a
few drops remaining in the vial.
Add one DPD No. 1 tablet straight from the foil and crush the tablet
using a clean stirring rod.
Add the water sample to the 10 ml mark.
Close the vial tightly with the cap and swirl gently several times until
the tablet is dissolved.
Place the vial in the sample chamber making sure that the
marks
are aligned.
Press the [ZERO/TEST] key.
The method symbol flashes for approx. 3 seconds.
The result 1 is shown in the display.
mg/l Chlorine dioxide = result 1 x 1.9
b) in presence of Chlorine
Fill a clean vial with 10 ml of water sample.
Add one GLYCINE tablet straight from the foil to the water sample
and crush the tablet using a clean stirring rod.
Close the vial tightly with the cap and swirl gently several times until
the tablet is dissolved.
Fill a second clean vial (24 mm Ø) with 10 ml of the water sample
and perform zero calibration (see “Operation”).
Remove the vial from the sample chamber and empty the vial.
Add one DPD No. 1 tablet straight from the foil and crush the tablet
using a clean stirring rod.
Transfer the contents of the first vial (Glycine solution) into the pre-
pared vail.
Close the vial tightly with the cap and swirl gently several times until
the tablet is dissolved.
0.0.0
Zero
Test
CL6
CL6
0.0.0

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Lovibond MD100 and is the answer not in the manual?

Lovibond MD100 Specifications

General IconGeneral
DisplayLCD
TypePortable Photometer
Wavelength Accuracy± 1 nm
Operating Temperature0 – 40 °C
Auto - OFFYes
Measurement RangeVaries depending on parameter
AccuracyVaries depending on parameter
Light SourceLED
Sample Volume10 ml
ParametersMultiple (depends on the model)
OpticsHigh-quality interference filters

Related product manuals