Model 3000 Survey meter Users Manual Section 2
Ludlum Measurements, Inc. Page 2-11 April 2021
Note:
Tapping the LOG button will start a count and automatically
log the result. A long press of the LOG button during the
count will toggle the display between the timer and reading.
Audio ‘clicks’ are disabled in COUNT mode by default
Device Count Audio Mode will enable\disable click audio
during a COUNT.
0 Disabled (Mute icon will always be illuminated in
COUNT mode.)
1 Enabled
Note:
Sigma audio is disabled in Count Mode. If Sigma Audio and
Count Mode Audio are enabled, click audio will be utilized in
Count Mode. If a result in terms of activity is desired, the
scaler count can also be in units of "d" or disintegrations. But
if the count units are chosen to be cpm or cps, then the
resulting answer is an averaged count rate over the time
interval. Similarly, if count units of Bq or dpm are chosen, the
resulting answer is an averaged disintegration rate.
Note:
If the user desires the instrument to show results in terms of
disintegrations/area (eg. dpm/100cm
2
or Bq/cm
2
), then the
appropriate factor should be placed in the Efficiency
parameter.