Model 3000 Survey meter Users Manual Section 4
Ludlum Measurements, Inc. Page 4-5 April 2021
Bypass Alarms: All Parameters except for Rate and Count Alarms
are protected from change through the Setup menu.
Bypass Alarms Count: All Parameters except for Rate Alarms,
Count Alarms, and Count Time are protected from change through
the Setup menu.
These options can be set through the Lumic Calibration Software. Locate
the Parameter “Device Setup Prtct” in the Lumic Calibration software
Others tab and enter the desired option into the “Value” field. By default,
when DIP Switch 2 is positioned in the ON position, Device Setup Prtct is
set to "Normal" operation.
In all cases in Setup Protect mode in which a Parameter is protected from
change, the value may still be viewed but may not be changed.
DISPLAY BACKLIGHT ‘Continuous On’: The Model 3000 display
backlight can be set to remain on continuously during operation. Follow the
steps above for SETUP PROTECT, but use DIP Switch 1 for display
backlight selection. Setting DIP Switch 1 to the ON (forward) position will
configure the display backlight to remain on during operation. Set DIP Switch 1
to the OFF (back) position, and the display will be backlit only when light levels
are low.
Note:
Setting the display backlight for continuous-on operation can
result in reduced battery life. The backlight can be configured to
always be off as well through the device menu or Lumic
Calibration Software.
Setup Mode Operation
Setup Page Selection: Once the Model 3000 is in Setup mode, the Setup Page
selection will be displayed on the LCD, and the Page number will be blinking,
indicating it as the selected item. The number of available parameters per Setup
Page will be displayed using the graph-the number of segments indicating the
number of parameters. Tap the MODE button to choose the Setup Page. Once
the desired Setup Page is shown, tap the UNITS button to move to the first
parameter of that Setup Page. Long-pressing the UNITS button will exit the
menu.