Model 3000 Survey meter Users Manual Section 4
Ludlum Measurements, Inc. Page 4-6 April 2021
Figure 11: Setup Page Selection display (showing page 4).
Long-pressing the UNITS button, no matter what parameter is shown, will first
return the user back to the Page Selection screen and then exit if held for
approximately 5 seconds in total. Exiting can also be accomplished by one or
two individual long presses (at least a half second) of the UNITS button
depending on where the user is in the menu. (If on the page selection, only one
press will exit, but if viewing an individual parameter, two presses will be
required.).
The list on the following page includes the five setup pages and parameters, in
order, on each page.
Model 3000 List of Parameters (in order)
Page 1 (P-1)
Calibration Constant Mantissa
Calibration Constant Exponent
Dead Time Correction 1
Dead Time Correction 2 Mantissa
Dead Time Correction 2 Exponent
Efficiency
High Voltage Setting
Pulse Threshold
Detector Current Overload Threshold
Loss of Count Alarm Time
Page 2 (P-2)
Primary Units and Minimum Display
Primary Units Maximum Display