Model 3000 Survey meter User’s Manual Section 7
Ludlum Measurements, Inc. Page 7-2 April 2021
Units
Depending on the chosen display units, different features will affect value of
the reading. The following table lists the features that apply to each of the
display units.
R to Sv Conversion
The R to Sv conversion is a setting available in Lumic Calibration software. It
defines the conversion factor between R and Sv. Since the calibration
constant is directly tied to R (counts per R), calibrating Sv requires a correct
setting of both calibration constant and R to Sv conversion.
Example: An R to Sv conversion factor of 0.0106 will cause a reading of
10.6 mSv/h while in a 1 R/h field.
Dead time correction
Efficiency
Dead time correction
Calibration constant
Dead time correction
Calibration constant
R to Sv conversion
These apply to all modes.