Model 3000 Survey meter Users Manual Section 5
Ludlum Measurements, Inc. Page 5-3 April 2021
Datalogging Operation Mode 2
Datalogging Mode 2 will allow the user to choose the Location ID (by Location
ID Table index) to store with the logged data.
When the LOG button is pressed, the current radiation readings and
other log data are temporarily stored.
In COUNT Mode: (Datalogging Mode 1 is always utilized in COUNT
Mode.)
- With a set count time, the LOG button is disabled until completion
of a Scaler count.
- For a continuous Scaler count (Scaler time is 0), the LOG button is
enabled at all times.
The LCD display will show a possible Location ID Table index for the
user. The index will be auto-incremented from the previously used
index.
Figure 13: Display showing a Location ID Table index of 36.
The ones digit of the index will be blinking, indicating a changeable
value. The user may then enter the preferred Location ID Table index
by tapping the ON/OFF (to select the digit) and MODE (to change
the value) buttons as in Setup mode.
Once the user has the preferred Location ID Table index entered, tap
the LOG button to save the log data.
For approximately 2 seconds, the LCD will display the Datalog Table
index for the newly saved log data. See Figure 12.
After displaying the Datalog Table index, the instrument will return to
the previous mode of operation.
The Datalog Table is linear; once the table is full, no further writes will
be allowed until the Datalog Table is erased.