Model 3000 Survey meter Users Manual Section 3
Ludlum Measurements, Inc. Page 3-2 April 2021
Backlight: built-in ambient light sensor automatically activates low-power
LED backlight, unless internal dip switch is set to continuous-on (will reduce
battery life)
User Controls:
ON/OFF/ACK long press to turn ON, tap to acknowledge alarms
and silence alarm tone, long press to reset Sigma Audio alarm, long
press to reset Rate, hold for OFF
MODE alternates between RATE (count rate), MAX ( captures peak
rate), and COUNT ( user-selectable preset count time from 0 to 10
minutes)
AUDIO turn “clickaudio on/off, turn Sigma Audio beep on/off
UNITS changes the units from count rate (cpm, cps) to
dose/exposure (µSv/h, mR/h) or disintegration (dpm/Bq)
Response Time: user-selectable from 1 to 60 seconds, or auto-response rate
FAST or SLOW
Audio click audio greater than 65 dB at 0.6 m (2 ft), alarm audio at greater than
72 dB, approximately 4.5 kHz
Power: four alkaline “AA” batteries
Battery Life: approximately 750 hours of operation (as low as 100 hours with
backlight configured for continuous-on), 16-hour low-battery warning
Maximum Current: 35 mAdc
Construction: high-impact plastic with water-resistant rubber seals and
separate battery compartment
Temperature Range: -20 to 50 ˚C (-5 to 122 ˚F), may be certified for
operation from -40 to 65 ˚C (-40 to 150 ˚F)
Environmental Rating: NEMA (National Electrical Manufacturers
Association) rating of 5 or IP (Ingress Protection) rating of 53
Size: 16.5 x 11.4 x 21.6 cm (6.5 x 4.5 x 8.5 in.) (H x W x L)
Weight: 0.9 kg (2.0 lb)