Model 3000 Survey meter Users Manual Appendix A
Ludlum Measurements, Inc. A-6 April 2021
Table A-2: AuxComm Mode
Setting
Mode
Baud Rate
0
SLURM
115200
1
LMI Direct
115200
2
375 Standard
2400
3
375 Ethernet
2400
4
375 for 272D
2400
SLURM: Used to encapsulate “LMI Direct” messages with encryption and checksums. This
mode is primarily used for Bluetooth®.
LMI Direct: Full exposure to the “LMI Comm” Communication protocol. This allows
reading and writing all parameters, data streaming, etc.
375 Standard: Legacy support for Model 375 Standard output. This is compatible with the
Model 375 Logger software and some versions of the Model 3276 Ethernet Kit (LMI#
4498-958).
375 Ethernet: Extended version of “375 Standard” that is intended for use with Model 3276
Ethernet Kit (LMI# 4498-958 **FW# N09 Only**) and the Ludlum Universal Software.
375 for 272D: Special version of “375 Standard” that makes the 272D display follow the
instrument display more closely with regards to leading zeros and decimal places.
Note:
The Model 3276 is the only unit that is wired to properly interface with
the 375 Ethernet module.
AuxComm Setting - Auto Off Time
This setting defines the number of minutes the AuxComm port will sit idle before turning
off. The primary use for this parameter is to save power while using Bluetooth® by turning
off the Bluetooth® module if there is no connection. Table A-3 shows valid settings.
Note:
The Model 3276 is the only unit that is wired to properly interface with
the 375 Ethernet module.
Table A-3: AuxComm Auto Off Time