Model 3000 Survey meter User’s Manual Section 1
Ludlum Measurements, Inc. Page 1-2 April 2021
Setup of the instrument is accomplished through the front-panel
buttons, or through software available from Ludlum Measurements.
The advanced user or administrator can set:
Calibration Constant
Dead Time Correction
Dead Time Correction 2
Efficiency
High Voltage
Pulse Threshold
Detector Current Overload Threshold
Loss of Count Alarm Time
Primary and Secondary Units
Primary and Secondary Minimum and Maximum Displays
Primary and Secondary Unit Alarm Levels
Primary and Secondary Count Units
Primary and Secondary Count Units Minimum Displays
Primary and Secondary Count Alarm Levels
Response Time
Auto-Response Rate (Fast or Slow)
Available Operational Modes
Count Time
Auto Shutdown Time
Backlight Threshold
Sigma or Click Audio Mode
Dead time correction coefficients can be configured either through the
device setup menu or through a wizard included in the calibration
software. Dead time correction employs first and second order
corrections for extended performance.
The front-panel setup feature can be disabled via the internal switch on
the Model 3000 in order to protect settings from inadvertent changes.
The unit is operated with four alkaline AA batteries for operation from
-20 to 50 °C (-5 to 122 °F). Battery life is approximately 750 hours
under normal usage. A low-battery indicator on the LCD warns when
less than 16 hours of battery life remain.
The Model 3000 may optionally be provided with a 1/8-inch stereo
headphone jack that would allow for headphone use for audio output. It
is located to the left of the detector connector.