ISR 12-LO • IGAR 12-LO Operating Manual Data format UPP (Universal Pyrometer Protocol) • 36
X = 0 ... 8 (dec.) 0 = Maximum value storage off
1 = 0.01 s 4 = 1.00 s 7 = external deletion
2 = 0.05 s 5 = 5.00 s 8 = automatically
deletion
3 = 0.25 s 6 = 25.00 s 9 = hold
Simulation of an external deletion contact
X = 0 metal mode from channel 1 and 2
X = 0 or 1 0 = 0 to 20 mA 1 = 4 to 20 mA
Output: XXXXYYYY (hex 8-digit, °C or °F)
XXXX = beginning of temperature range
YYYY = end of temperature range
Reading temperature
sub range:
XXXX (hex. 4-digit) beginning of temp. range
YYYY (hex. 4-digit) end of temp. range (°C or °F)
00 ... 97 = regular device addresses
99 = global address with response
98 = global address without response (only setting
1 = 2400 baud 4 = 19200 baud (7 is not allowed)
2 = 4800 baud 5 = 38400 baud 8 = 115200 baud
3 = 9600 baud 6 = 57600 baud
Output: X = 0: display in °C; X = 1: display in °F
Output: XXXX, XXXX = 0000 ... 9999 in ‰
XX = 00 ... 99% (2 digit, hex.)
XX=00 … FF (00 = no error)
(01 … FF: error code for LumaSense service)
Output: XXX (dec. 000 ... 099 °C or 032 ... 210 °F)
Max. internal
temperature:
Output: XXX (dec. 000 ... 099 °C or 032 ... 210 °F)
Output: X = 1 or 2; 1 = RS232, 2 = RS485
X = 0 switch off laser targeting light
X = 1 switch on laser targeting light
1 = lock lk1, removal with command lk0 or power
off-on
0 = removal of lock lk1
3 = continuous lock lk3, removal only with command
lk2