ISR 12-LO IGAR 12-LO Operating Manual Data format UPP (Universal Pyrometer Protocol) 37
Description
Command
Parameters
Reading parameters: AApa
Output decimal 11-digit:
Digit 1 und 2 (10...99 or 00): (emissivity)
Digit 3 (0 ... 6): t
90
(exposure time)
Digit 4 (0 ... 8): t
CL
(clear time of max.
storage)
Digit 5 (0 ... 1): analog output
Digit 6
und 7: (00 ... 98): pyrometer’s
temperature
Digit 8 und 9 (00 ... 97): address
Digit 10 (0 ... 6 or 8): baud rate
Digit 11 (0): (0 or 1): keyboard active or
locked
Digit 12 ... 15 (0800 ... 1200): emissivity slope
Autoreset: AAre Activating auto-reset, initializes the pyrometer new
and resets the instrument to default settings
Reference number:
AAbn
Output: XXXXXX (hex 6-digit)
Serial number:
AAsn
Output: XXXXX (5-digit, decimal)
Device type /
software version:
AAve Output: XXYYZZ (6-digit, decimal)
XX = 63 (ISR 12-LO or IGAR 12-LO)
YY = month of software version
ZZ = year of software version
Software version
in detail:
AAvs tt.mm.yy XX.YY
tt = day; mm = month; yy = year; XX.YY = software
version
Device type: AAna Output: “ISR 12-LO“ or “IGAR 12-LO
(16 ASCII-characters)
Note: The letter “l” means the lower case letter of “L”.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the LumaSense technologies IGAR 12-LO and is the answer not in the manual?

LumaSense technologies IGAR 12-LO Specifications

General IconGeneral
BrandLumaSense technologies
ModelIGAR 12-LO
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals