IS 12 / IGA 12 Manual Controls and Installation 19
Warning: To reduce the risk of injury to the eyes, do not look
directly into the targeting laser and do not point the targeting laser
into anyone's eyes. The instrument is equipped with a class II laser
that emits radiation.
Note: The pyrometer can measure objects at any distance but the object has to be
bigger than or at least as big as the spot size of the pyrometer in the measuring
distance.
Calculating the spot sizes on different measuring distances:
Spot sizes for other measuring distances can
be calculated with the following equations
or with the IR calculator of the InfraWin
software.
Table values: a
1
= measuring distance
M
1
= spot size
D = aperture
3.3.2 Focusable optics
Focusable optics IS 12
Measuring
distance
a [mm]
Spot size M
90
[mm]
Objective
length
S [mm]
MB 14
MB 16
MB 18; 20L
MB 25; 35L
Optics
1
a = 277 mm
0.9
0.6
0.6
0.4
30
a = 400 mm
1
0.8
0.8
0.5
9.5
a = 533 mm
1.4
1.1
1.1
0.7
0
Optics
2
a = 388 mm
1
0.8
0.8
0.45
30
a = 700 mm
1.8
1.5
1.5
0.8
8.6
a = 1170 mm
3
2.4
2.4
1.4
0
Optics
3
a = 550 mm
1.5
1
1
0.6
30
a = 3000 mm
9
6
6
3.3
3.3
a = 9500 mm
25
19
19
10.6
0
Aperture D [mm]*
13.5 to 17
13.5 to 17
10 to 13
5 to 7

Table of Contents