C - Temperature unit is
F - Temperature unit is
2) After Display unit is selected to " C " or " F ", press the
" Enter Button " ( 3-8, Fig. 1 ) will save the setting
function with default.
10-8 Set CD temperature compensation factor
When the lower display show " PCNT "
1) This function only for the Conductivity ( PCNT ) mode of
adjusting the probe's Temp.compensation value in
%/per unit. The default value is 2 %/ per .
2) Use the " Button " ( 3-5, Fig. 1 ) or " Button "
( 3-6, Fig. 1 ) to select the upper value to the desired
Temp. compensation value ( %/per ), then press the
" Enter Button " ( 3-8, Fig. 1 ) will save the setting
value temporally.
10-9 CELL K Value setting
When the display show " CELL "
When the Display show the text " CELL ", Use the
" Button " ( 3-5, Fig. 1 ) or " Button " ( 3-6, Fig. 1 ) Adjustment
CELL(Probe) K Value setting, than press the " Enter Button "
( 3-8, Fig. 1 ) will save the setting.
Remark :
During execute the " Advanced Setting " function,
if press " HOLD Button " ( 3-3, Fig. 1 ) will exit the
" Advanced Setting " function, the LCD will return
to normal screen.