EasyManua.ls Logo

Lutron Electronics WA-2015 - 7-2 Calibration

Lutron Electronics WA-2015
49 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1)Prepare the Oxygen Probe ( optional, DOPB-11 ),
install the " Probe Plug " ( 6-1, Fig. 4 ) into the
" DO Socket " ( 3-13, Fig. 1 ).
2)Power on the meter by pressing " Power Button "
( 3-2, Fig. 1 ) once.
Select the measurement mode to " Dissolved Oxygen
measurement ". refer chapter 4, page 11.
The Display will show " mg/L " and " Temp. " indicator.
Calibration at first !
Press the " Function Button " ( 3-6, Fig. 1 ) once,
the Display will show the " % O2 " indicator.
Wait approx. 2 minute until the reading value
reach stable, if the reading value not within
20.7 to 21.1 ( 20.9 ±0.2 ), then should be
processed the calibration procedures first.
The calibration procedures, please refer
chapter 7-2, page 28.
After execute the calibration procedures, the
display should show the value near 20.9
( 20.8 to 21.0 ).
3)Press the " Function Button " ( 3-6, Fig. 7 ) once,
the right low display will show the " mg/L " unit,
the meter is ready for the Dissolved Oxygen
measurement.
4)a.Immersed the probe to a depth at least 10 cm of
the measured liquid in order for the probe to be
influenced by the temperature & automatic
temperature compensation to take place.
26

Related product manuals