6. CONDUCTIVITY/TDS MEASURING
and CALIBRATION PROCEDURE
The meter default function are following :
*The display unit is set to conductivity ( uS, mS ).
*The temperature unit is set to .
*Temp. compensation factor is set to 2.0% per C.
*Auto range.
*Auto power off.
*The sampling time of data logger function is
2 seconds.
Fig. 3
If the meter is first time to connect
the conductivity probe, it should
execute the calibration procedures
before operation, the calibration
procedures refer chapter 6-3, page 22.