Operation Manual easyTymp
8111254 Rev. 9
59
17/01/2018
Guidance and manufacturer’s declaration — electromagnetic immunity
The easyTymp is intended for use in the electromagnetic environment specified below. The customer or the user of
the device should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic environment guidance
Portable and mobile RF communications equipment should be
used no closer to any parts of the device, including cables,
than the recommended separation distance calculated from
the equation applicable to the frequency of the transmitter.
Recommended separation distance:
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to
80 MHz
3 Vrms
Pd 2,1
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to
2,5 GHz
3 V/m
Pd 2,1
80 MHz to 800 MHz
Pd 3,2
800 MHz to 2.5 GHz
Where P is the maximum output power rating of the transmitter
in watts (W) according to the transmitter manufacturer and d is
the recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by
an electromagnetic site survey
(a)
, should be less than the
compliance level in each frequency range
(b)
.
Interference may occur in the vicinity of equipment marked
with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
(a)
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey
should be considered. If the measured field strength in the location in which the device is used exceeds the
applicable RF compliance level above, the device should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as reorienting or relocating the device.
(b)
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.

Other manuals for Maico easyTymp

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Maico easyTymp and is the answer not in the manual?

Maico easyTymp Specifications

General IconGeneral
BrandMaico
ModeleasyTymp
CategoryMedical Equipment
LanguageEnglish

Related product manuals