Operation Manual easyTymp
6.6 Test Protocols
NOTE: Test protocols are configuration dependent.
Tympanometry, Frequency: 226 Hz
Earside: Ipsilateral
Tympanometry, Frequency: 1 kHz
Earside: Ipsilateral
03 226Hz + Ipsi Reflex Auto
Tympanometry, Frequency: 226 Hz
Number of Reflexes tested = 4, Frequencies: 0.5, 1.0, 2.0, 4.0 kHz
Intensity Reflex Min (Intensity in dB HL) = 70
Intensity Reflex Max (Intensity in dB HL) = 100
Probe frequency during reflexes: 226 Hz
Earside: Ipsilateral
04 226Hz + Ipsi Reflex 90 dB
Tympanometry, Frequency: 226 Hz
Number of Reflexes tested = 4, Frequencies: 0.5, 1.0, 2.0, 4.0 kHz
Intensity Reflex (Intensity in dB HL) = 90
Probe frequency during reflexes: 226 Hz
Earside: Ipsilateral
05 1kHz + Ipsi Reflex Auto
Tympanometry, Frequency: 1 kHz
Number of Reflexes tested = 4, Frequencies: 0.5, 1.0, 2.0, 4.0 kHz
Intensity Reflex Min (Intensity in dB HL) = 70
Intensity Reflex Max (Intensity in dB HL) = 100
Probe frequency during reflexes: 226 Hz
Earside: Ipsilateral
06 1kHz + Ipsi Reflex 80 dB BB
Tympanometry, Frequency: 1 kHz
Number of Reflexes tested = 1, Test signal: Broad-band noise
Intensity Reflex (Intensity in dB HL) = 80 dB
Probe frequency during reflexes: 226 Hz
Earside: Ipsilateral
07 226Hz + Ipsi-Contra Auto
Tympanometry, Frequency: 226 Hz
Number of Reflexes tested = 8, Frequencies: 0.5, 1.0, 2.0, 4.0 kHz
Intensity Reflex Min (Intensity in dB HL) = 70
Intensity Reflex Max (Intensity in dB HL) = 100
Probe frequency during reflexes: 226 Hz
Earside: Ipsi- and Contralateral