mptying the Waste Bottle
Use this procedure to empty the waste bottle.
1. Verify that the system is in an idle state.
2. Unscrew the grey cap of waste bottle by turning the lid counter-clockwise.
3. Empty the waste according to your laboratory waste handling procedures.
4. Reattach the cap by turning it clockwise until it is snug.
Wash Procedures
Use the following procedures as required.
C
ell and Syringe Wash
Use this procedure to perform a quick wash of the cell and the syringe to prepare for an
experiment.
Fluid requirement: Buffer, water, and methanol
Time requirement: 9.4 minutes
P
rocedure:
1. On the Instrument Controls tab, click Cell and Syringe Wash.
2. Follow the instructions on the screen.
Syringe Fill
Use this procedure to fill the syringe with titrant before an experiment.
Fluid requirement: titrant in microcentifuge tube
Time requirement: 2.3 minutes
P
rocedure:
1. On the Instrument Controls tab, click Syringe Fill.
2. Follow the instructions on the screen.