EasyManua.ls Logo

MAPADA V-1600 - Page 14

MAPADA V-1600
36 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1600/1800 Manual
13
press the number ahead of the unit directly by the numeric keypad.
If you don’t want to select any item, press
RETURN
RETURNRETURN
RETURN
to go back to last interface.
Set Quantity of Standard Samples
Move the cursor on Number” by pressing
and
, followed by
pressing
ENTER
. (Fig. 14)
ENTER
RETURN
Fig. 14
Input the number of your standard samples by numeric keypad. After
pressing
ENTER
,it returns to the setting interface.
Notethe number range is 1-9other number is invalid.
Set Concentration of Standards
Before this step, you must blank first. Pull the blank cuvette in the light
path in the pre-testing interface, press
ZERO
.
Move the cursor in Concentration”, following press
ENTER
.(Fig. 15)
ENTER
RETURN
Fig.15
Pull sample 1 in the light path, input the concentration of sample 1 by
numeric keypads, then press ENTER. It goes into the next setting
interface of sample 2, the operation is the same as sample 1. Repeat the
operation one by one till the last sample finished, and it goes back to the
setting interface.
Standard Curve
Unit
Number
Standard Curve
N=2
Please Input New
N=
Pull Sample 1
C=1.500
Please Input New
C=
Standard Curve
Unit
DOMINIQUE DUTSCHER SAS