SWEEP
OPERATION
IMODE
Data type :
Allowed suffices
:
Default suffix
:
Example:
SWEEP
:MKRON
:MKROFF
Data
type :
Allowed suffices
:
Default suffix
:
Examples:
:CFRQ
:<cmd>
:RFLV
:<cmd>
:LFGF
:<cmd>
:LFGL
:<cmd>
:INTF
:<cmd>
Select
Instrument Mode
Character
Program Data (either NORMAL, NOISE1
OR
NOISE2 for signal generator operation
or SWEEPER for swept
operation
)
None
None
IMODE SWEEPER
[not used alone]
Enable Sweep
Markers
Disable sweep Markers
None
None
None
SWEEP
:CFRQ: START 75MHZ;STOP 150MHZ;STEP
100; TIME 10MS
SWEEP
:RFLV: START -56DBM;STOP -12DBM;STEP
440;TIME 25MS
SWEEP
:CFRQ: MKRNUM
1?
VALUE 83MHZ
;MKRON
[not used alone]
Select
Carrier
Frequency
sweep
parameter entry where <cmd>
is replaced
by one of the
commands
(START,
STOP, STEP.
TIME, MKRNUM, MKRON, MKROFF
or
VALUE)
[not used alone]
Select RF
Level sweep parameter entry where
<cmd> is
replaced
by one
of the
commands
(START, STOP. STEP,
TIME, MKRNUM,
MKRON, MKROFF or VALUE)
[not used
alone]
Select LF Generator Frequency
sweep
parameter
entry where
<cmd> is
replaced by one of the commands (START,
STOP,
STEP, TIME, MKRNUM,
MKRON, MKROFF or VALUE)
[not used alone]
Select LF Generator
Level sweep parameter entry where
<cmd> is
replaced by one of the commands
(START. STOP,
STEP, TIME,
MKRNUM, MKRON, MKROFF or VALUE)
[not used alone]
Select
Internal Modulation Oscillator
Frequency sweep
parameter entry where
<cmd> is replaced by one of the
commands
(START, STOP, STEP. TIME,
MKRNUM,
MKRON, MKROFF or VALUE)
3-2-32