METROLOGICAL DETAILS
Class III
Number of verifi cation scale intervals ≤ 6000
Load cell excitation power supply (V DC) 5 V DC
Maximum signal voltage for dead load (mV) 10 mV
Minimum signal voltage for dead load (mV) 0,05 mV
Minimum input-voltage per verifi cation scale interval, e (μV) 0,03 µV/e
Measuring range minimum voltage (mV) 6,05 mV
Measuring range maximum voltage (mV) 20 mV
Operating temperature range (°C) 0-40 ºC
Minimum load cell impedance 44 Ω
Maximum load cell impedance 1050 Ω
Value of the fractional error pi 0.5
Recommended specifi cation of the load cell cable:
Type 6 x 0.5 mm2
Size < 100 m
Section ≤ 0.5 mm2
Impedance < 5.4 Ω
ENERGY
Internal Commuter Source (AC version) 100-240 V AC
Min power 15 W
Frequency 50-60 Hz
External adaptor (DC version) 12 V DC
Min Power 12 W
Conector 2.1x5.5x10 mm
Polarity (positive in the centre)
COMMUNICATIONS
RS-232 Series (TX/RX) Up to 2 ports (not possible with extra connections, like Ethernet)
Ethernet (10/100) 1 (in case the model has this connection)
WIFI (2.4GHz) 1 (in case the model has this connection)