Martin maxium & maxium Beamer Operating Instructions
V 2.0
78
10.10.3 Technical Data
Power supply: through the maxium HF unit
Device fuse: M1.6A
Power input/requirements: max. 40 W
Protection class: I
Classification acc. to MDD: II b
Argon gas flow, CUT:
0.1 to 12 l/min ± 20%; can be switched off by setting a rate of 0.0 l/min
Argon gas flow, COAG:
0.1 to 12 l/min ± 20%; can be switched off by setting a rate of 0.0 l/min
LF leakage currents: acc. to IEC 601, Part 1 (checked together with maxium HF generator)
Type of applied part: CF; defibrillation-proof
Mode of operation: INT 10s/30s
Weight: 3.5 kg
Interference suppression: Limit values acc. to EN 55011 and IEC 601-1-2
Interference immunity acc. to IEC 1000
Dimensions: Width 390 mm
Height 108 mm (without ball-head bolts)
Depth 435 mm
Environmental conditions for
transportation and storage
Ambient temperature
Relative humidity
Atmospheric pressure
-25°C to +70°C
10% to 100%
500 hPa to 1060 hPa
Environmental conditions
for operation
Ambient temperature
Relative humidity
Atmospheric pressure
+10°C to +40°C
30% to 75%
700 hPa to 1060 hPa
Conforms to Directive 93/42/EEC