Martin maxium & maxium Beamer Operating Instructions
V 2.0
87
14.1 Technical Specifications & Certificate
Power requirements 100–240 V ± 10%; 50/60 Hz
Line current max. 6.3 A
Line fuse 6.3 A
Nominal frequencies 300 / 400 / 600 kHz
Power input without HF output 40 VA
at max. power output 600 VA
Output power maximum cutting power 400 W
maximum coagulation power 320 W
Current Type HF Power Crest Facto
HF Voltage
Pure Cut 5–400 W at 400 ohms CF 1.5 max. 1,200 V
p
max. 2,400 V
pp
Blend Cut 5–320 W at 400 ohms CF 1.9 max. 1,400 V
p
max. 2,800 V
pp
Super Blend 5–250 W at 400 ohms CF 2.4 max. 1,400 V
p
max. 2,800 V
pp
Uro Cut Pure 5–400 W at 600 ohms CF 1.5 max. 1,600 V
p
max. 3,200 V
pp
Uro Cut Blend 5–320 W at 600 ohms CF 1.9 max. 1,600 V
p
max. 3,200 V
pp
Argon Cut 5–400 W at 400 ohms CF 1.5 max. 1,200 V
p
max. 2,400 V
pp
Polyp I* 1–120 W at 150 ohms n. a. max. 1,200 V
p
max. 2,400 V
pp
Polyp II* 1–120 W at 150 ohms n. a. max. 1,200 V
p
max. 2,400 V
pp
Papilla I** 1–50 W at 400 ohms n. a. max. 2,000 V
p
max. 4,000 V
pp
Papilla II* 1–100 W at 150 ohms n. a. max. 1,500 V
p
max. 3,000 V
pp
Care Cut 1–70 W at 600 ohms CF 1.7 max. 750 V
p
max. 1,500 V
pp
Clamp Coag 5–250 W at 250 ohms CF 3.0 max. 1,500 V
p
max. 2,800 V
pp
Forced Clamp 5–150 W at 300 ohms CF 4.6 max. 2,000 V
p
max. 4,000 V
pp
Contact Coag 5–250 W at 100 ohms CF 3.3 max. 2,000 V
p
max. 4,000 V
pp
Contact Coag Auto Stop 5–250 W at 100 ohms CF 3.3 max. 2,000 V
p
max. 4,000 V
pp
Forced Coag 5–150 W at 300 ohms CF 5.4 max. 2,800 V
p
max. 5,000 V
pp
Uro Coag 5–150 W at 600 ohms CF 5.7–4.4 max. 3,700 V
p
max. 6,000 V
pp
Forced Uro Coag 5–150 W at 600 ohms CF 5.7 max. 4,500 V
p
max. 7,500 V
pp
Macro Spray 2–150 W at 600 ohms CF 20–9.0 max. 8,000 V
p
max. 11,000 V
pp
Spray 2–150 W at 600 ohms CF 9.0 max. 8,000 V
p
max. 11,000 V
pp
Endo Spray 1–100 W at 500 ohms CF 18–5.4 max. 4,000 V
p
max. 5,900 V
pp
Duo Spray 2–150 W at 600 ohms CF 9.0 max. 8,000 V
p
max. 11,000 V
pp
Care Coag 1–70 W at 100 ohms CF 1.7 max. 250 V
p
max. 500 V
pp
Care Coag Auto Stop 1–70 W at 100 ohms CF 1.7 max. 250 V
p
max. 500 V
pp
Desiccation 1–120 W at 100 ohms CF 1.7 max. 350 V
p
max. 700 V
pp
Desiccation Auto Stop 1–120 W at 100 ohms CF 1.7 max. 350 V
p
max. 700 V
pp
Argon Beam 2–150 W at 600 ohms CF 9.0 max. 8,000 V
p
max. 11,000 V
pp
Argon Endo 1–100 W at 500 ohms CF 18–5.4 max. 4,000 V
p
max. 5,900 V
pp