Martin maxium & maxium Beamer Operating Instructions
V 2.0
88
Current Type HF Power Crest Facto
r
HF Voltage
Pulsed Argon: slow repetition rate*
1–100 W at 500 ohms (CF 18–5.4) max. 4,000 V
p
max. 5,900 V
pp
Pulsed Argon: fast repetition rate *
1–100 W at 500 ohms (CF 18–5.4) max. 4,000 V
p
max. 5,900 V
pp
Bipo Pure Cut 1–120 W at 300 ohms CF 1.7 max. 600 V
p
max. 1,200 V
pp
Bipo Blend Cut 1–120 W at 300 ohms CF 2.1 max. 850 V
p
max. 1,700 V
pp
Forfex 1–120 W at 50 ohms CF 1.9 max. 180 V
p
max. 360 V
pp
Micro Coag 0.1–50 W at 200 ohms CF 1.8 max. 200 V
p
max. 400 V
pp
Micro Stop 0.1–50 W at 200 ohms CF 1.8 max. 200 V
p
max. 400 V
pp
Auto Micro 0.1–50 W at 200 ohms CF 1.8 max. 200 V
p
max. 400 V
pp
Macro Coag 1–120 W at 50 ohms CF 1.9 max. 180 V
p
max. 360 V
pp
Macro Stop 1–120 W at 50 ohms CF 1.9 max. 180 V
p
max. 360 V
pp
Auto Macro 1–120 W at 50 ohms CF 1.9 max. 180 V
p
max. 360 V
pp
Seal Safe 5–320 W at 25 ohms CF 1.9 max. 300 V
p
max. 600 V
pp
Endo Seal 1–100 W at 50 ohms CF 2.1 max. 230 V
p
max. 450 V
pp
Note: V
p
= peak voltage; V
pp
= peak to peak
* Data applies to coagulation phase only; cycle-based average value deviates.
** Data averaged across cycle; pulse-based average value deviates.
n.a. = not applicable
Only the following currents are permitted for use with the baby NE (maximum power indicated in each case):
Pure Cut 100 W Contact Coag 90 W Argon Beam 100 W Clamp Coag 90 W
Blend Cut 100 W
Contact Coag Auto Stop
90 W
Argon Endo 100 W Forced Clamp 90 W
Super Blend 100 W Spray 80 W Care Coag 50 W
Pulsed Argon: slow
repetition rate 100 W
Argon Cut 100 W Endo Spray 100 W Care Coag Auto Stop 50 W
Pulsed Argon: fast
repetition rate 100 W
Care Cut 70 W