Protection class I acc. to IEC 601-1
Classified acc. to MDD II b
Leakage currents (LF/HF) in compliance with IEC 601,
Type (of applied part) CF; defibrillation-proof
Duty type intermittent 10 s / 30 s = duty factor of 25%
Device fuses 2 x T 6.3 A (slow-blow)
Sound level HF current activation: 55 dB(A) (adjustable 45–60 dB)
Alarm: 65 dB(A)
Weight 8.3 kg
Interference suppression Limits in conformity with EN 55011 and IEC 601-2-2
Interference immunity in conformity with IEC 1000
Dimensions Width 390 mm
Height 182 mm
Depth 435 mm
Environmental conditions
for transportation and stor-
age
Ambient temperature
Relative humidity
Atmospheric pressure
-25°C to +70°C
10% to 100%
500 hPa to 1060 hPa
Environmental conditions
for operation
Ambient temperature
Relative humidity
Atmospheric pressure
+10°C to +40°C
30% to 75%
700 hPa to 1060 hPa