Test Specifications for ENCLOSURE PORT IMMUNITY to RF Wireless
Communication Equipment
Test
Frequency
(MHz)
Band
(a)
(MHz)
Service (a) Modulation
(b)
Maximum
Power
(W)
Distance
(m)
Immunity
Test Level
(V/m)
385
380-
395
TETRA 400
Pulse
modulation
(b)
18 Hz
1.8 0.3 27
450
430-
470
GMRS 460, FRS 460
FM (c)
+/- 5 kHz
deviation
1 kHz sine
2 0.3 28
710
704-
787
LTE Band 13, 17
Pulse
modulation
(b)
217 Hz
0.2 0.3 9
745
780
810
800-
960
GSM 800/900, TETRA
800, iDEN 820, CDMA
850, LTE Band 5
Pulse
modulation
(b)
18 Hz
2 0.3 28
870
930
1720
1700-
1990
GSM 1800; CDMA
1900; GSM 1900;
DECT; LTE Band 1, 3.
4. 35: UMTS
Pulse
modulation
(b)
217 Hz
2 0.3 28
1845
1970
2450
2400-
2570
Bluetooth, WLAN,
802.11 b/g/n, RFID
2450, LTE Band 7
Pulse
modulation
(b)
217 Hz
2 0.3 28
5240
5100-
5800
WLAN 802.11 a/n
Pulse
modulation
(b)
217 Hz
0.2 0.3 9
5500
5785