MS8218 True RMS Multimeter
Technological Specifications
Range and Accuracy
The below-listed accuracies under different ranges refer to those which are guaranteed by the meter within one-year calibration,
with normal use under the operating temperature of ISoC-2SoC and relative humidity less than SO%.The presentation for
accuracy
is: ±
(**% reading digits
+
number of lower digits)
AC Voltage/DC Voltage + AC Voltage
- - -
Range Resolution
Accuracy
40Hz-1KHz 1KHz-10KHz 10KHz-20KHz
SOmV O.OOlmV
±(O.S% +40)
±(l
% + 40)
±(2.S%
+ 40)
SOOmV
O.OlmV
±(O.S% +40)
±(l
% + 40)
±(2.S%
+ 40)
SV O.lmV
±(O.S% +40)
±(l
% + 40)
±(2.S%
+ 40)
SOV
lmV
±(O.S% +40)
±(l
% + 40)
±(2.S%
+ 40)
SOOV
lOmV
±(O.S% +40)
±(l
% + 40)
unspecified
1000V O.lV
±(O.S% +40)
unspecified unspecified
Notes: above accuracies can be guaranteed within 10%-100% of the full range
.
DC Voltage
Range Resolution Accuracy
SOmV O.OOlmV
±
(0.03%+ 10)
SOOmV O.OlmV
±(0.03%+6)
SV O.lmV
±(0.03%+6)
SOV
lmV
±(0.03%+6)
SOOV
lOmV
±(0.03%+6)
1000V
o
.1V
±(0.03%+6)
Notes: above accuracies can be guaranteed
withinthe
full range