Chapter 4 Specication
CAUTION :
The Chiller needs to meet the requirements of the above table under the
conditions of a circulating temperature of 40 ° C and an outlet temperature of 22 ° C.
The above recommended water pressure requires the pressure drop of the main
line Δp ≤ 0.5bar. If this value is exceeded, the main circuit water pressure should be
increased accordingly.
Cooling water and lter element need to be replaced once a month; replace the
cooling water with glycol antifreeze every winter (below 0
environment). After the
0
environment is over, replace the antifreeze with pure water and replace the lter
to restore the maintenance frequency once a month.
4-
QBH
Water Cooling Condition
Cooling Method
Water pipe size
requirement
Water Flow
Rate
L/min
Hydraulic
pressure
bar
Cooling
Temperature
(℃)
Water cooling
Outer diameter * inner
diameter = Φ6 * Φ4
≥2 ≥3 28-30
NOTE:
External light path tube inner diameter ≥ 8mm, length ≤ 20m;
The length of the Φ6 pipe connected to the LOE after switching from the external
light path is ≤ 1m;
QBH is connected in series with the cutting head;
The above recommended external light path water pressure requires the pressure
drop of the cutting head Δp ≤ 1.5bar. If this value is exceeded, the external light path
water pressure should be increased accordingly.