Troubleshooting
42

The private alert feature allows adding an acoustic warning signal to the audio signal. This added
signal is audible only to the user of the speech processor and can be adjusted in 8 loudness
steps. Your audiologist will set the loudness accordingly.

If the battery voltage falls below a certain level, four short warning beeps will be generated
approximately every 14 seconds. You are still able to hear, but should change the batteries of
the OPUS 2 speech processor as soon as possible.

If a maximum or minimum value of volume or audio sensitivity has been reached, a continuous
beeping signal is audible for the user as long as the key of the FineTuner is pressed.

If a command from the FineTuner has been executed successfully by the OPUS 2 speech
processor, a confirmation beep is audible for the user of the speech processor.
The two warning signals and the confirmation signal may be deactivated permanently by your
audiologist if your prefer this.