14 | Preparation for use
6 Preparation for use
WARNINGS
For use only by medically trained persons who have been adequately trained in suction
procedures and in the use of aspirators.
CAUTIONS
– The Basic 30 must remain in an upright position during use.
– Sterile accessories must be checked on the integrity of the packaging before use.
– Non sterile and reusable accessories must be cleaned, disinfected and / or sterilized
according to the Medela cleaning guide (product code 200.2391).
– Check the Basic 30 system before use for damage of
the power cord or plug, obvious device damage or
safety defects and proper functioning of the device.
– Check for completeness and general condition of the
Basic 30 delivery package.
– Check all accessories prior to use:
– Suction jars, lids and liners for cracks, brittle and
flawed spots. Replace if necessary.
– Tubing for cracks, brittle areas and that connectors
are firmly attached. Replace if necessary.
– As an additional safety test, evacuate the system
(including jars) to maximum vacuum before actual
use.
Checks before use
6.1
3. Attach all necessary accessories according to your needs.
See chapters 17 / 18 for more details.
2. 2.1 Attach a filter (if required)
to the safety set with the
arrow pointing in the flow
direction. The filter must
be exchanged after an
overflow and after each
patient, but latest after a
week.
Assembly of the
basic configuration
1. 1.1 Make sure that the Safety
Set is attached to the
Basic 30. See chapter 5.5.
6.2