7 Operating instructions
CAUTIONS
The Basic 30 is to be set up in such a way, that a separation from the mains supply
can be easily managed.
Connect Basic 30
to mains power
Check the pump before use following the instruction in
chapter 6.1.
1.
7.1
2. 2.1 Plug in the mains plug of
the power cord to a fixed
mains socket.
3. 3.1 The LED on the housing
will light up to indicate
standby mode.
3. 3.1 Seal the end of the patient
tubing with your thumb.
3.2 Compare the maximum
vacuum according to the
specification (on left). See
chapter 11 if the maximum
vacuum is not reached.
+ 2000 m – 66 kPa
– 495 mmHg
+ 1000 m – 75 kPa
– 562 mmHg
+ 500 m – 80 kPa
– 600 mmHg
0 m – 85 kPa
– 638 mmHg
Check maximal
vacuum for
functional check
1. 1.1 Switch on the Basic 30.
The pump will start
running.
7.2
2. 2.1 Turn the vacuum regulator
to the right to set
maximum vacuum
Altitude
above
sea level:
(Tolerance: +/- 10%)
Specifications:
Max. Vacuum: