H4-EN Maintenance P
AGE
42/44
6
6.2 Maintenance planning
The table below shows the list of actions to be performed for a full preventive maintenance.
Each action must be performed by the user or a technician approved by Medisoft or its
representative.
The highlighted actions must be performed by the user by following the procedures
described in the next paragraphs. The procedures for the technicians are included in the
field service manual.
6.2.1 Maintenance planning for Ergocard CPX Clinical
Operation
Part
Frequency
Notes
Who performs?
Clean
Electronic module
1
year
User
Replace
Electrochemical O
2
cell
1 year
Technician
Umbilical
1 year
(1
)
User / Technician
Sample o
-
ring
1 year
Technician
Sample filter
1 year
Technician
Flow regulator
2 years
Technician
Check
Pneumatic circuit leak
1 year
Technician
Weather station
1 year
Technician
Analyse and calibration flow rate
1 year
Technician
Flow transducer offset and span
1 year
Technician
O
2
and CO
2
analysers offset and span
1 year
Technician
Test
Quality control tests
1 month
User
Biological control tests
1 month
User
Notes
(1) Sooner if the gas calibration fails due to a damaged umbilical.