H4-EN Maintenance P
AGE
43/44
6
6.2.2 Maintenance planning for Ergocard CPX Professional
Operation
Denomination
Frequency
Notes
Who performs?
Clean
Electronic
module
1
year
Technician
Replace
Inspiratory bag 9 liters
6 months
(1)
User
Inspiratory bag 40 liters
6 months
(2)
User
Umbilical
1 year
(
3
)
User / Technician
Sample o
-
ring
1 year
Technician
Sample filter
1 year
Technician
Flow
regulator
2 years
Technician
Demand valve or inspiratory bag o
-
ring
s
2 years
(
4
)
Technician
Check
Pneumatic circuit leak
1 year
Technician
Weather station
1 year
Technician
Analyzer
and calibration flow rate
1 year
Technician
Flow
transducer offset and span
1 year
Technician
O
2
and CO
2
analysers offset and span
1 year
Technician
Test
Quality control tests
1 month
User
Biological control tests
1 month
User
Notes
(1) Sooner if the inspiratory bag has a leak. Only for Ergocard CPX Professional with FRC-
N
2
inspiratory bag option.
(2) Sooner if the inspiratory bag has a leak. Only for Ergocard CPX Professional with
hypo-hyperoxia option.
(3) Sooner if the gas calibration fails due to a damaged umbilical.
(4) Only for Ergocard CPX Professional with FRC-N
2
option.