AVTM24-1J REV A April 2006
41
Section B
SAFETY
SAFETY IS THE RESPONSIBILITY OF THE USER
La seguridad es la responsabilidad del operador
F
WARNING
Do NOT connect this instrument to energized
circuits.
The line-operated 100-Ampere current source should be
used and maintained by trained personnel who are familiar
with the usual precautions for handling portable line-
operated equipment. For protection against electrical
shock the cabinet must be grounded, either by the line
cord or by a separate wire.
In using the instrument take care to avoid skin burns
caused by touching an overheated range module, undersize
bead wire, poor connections, or small item under test.
Overheating may occur if the tester is operated for more
than the rated duty cycle.