16 D24030 | Date 11/2018
Construction and description
2.11 Ambient conditions
► The appliance may only be used in enclosed rooms and under the following ambient
conditions:
Ambient temperature
16 ºC to 40 ºC
Humidity rh max. 70 %, non-condensing
Overvoltage category II
Pollution degree 2
Altitude of installation max. 2,000 m above sea level
► The appliance may not be used in areas where there is a risk of explosion. The ambient air
must not contain any explosive dusts, gases, vapours or gas-air mixtures. The appliance is
not explosion-proof.
► Heavy dust production or aggressive vapours in the vicinity of the appliance could lead to
sedimentation in the interior and, as a consequence, could result in short circuits or dam-
age to electrical parts. For this reason, sufficient measures to prevent large clouds of dust
or aggressive vapours from developing should be taken.
2.12 Scope of delivery
► Power cable
► Tilt protection
► Sliding grid (load capacity 30 kg each)
► USB storage medium with software and AtmoCONTROL manual
► The operating instructions at hand
► Calibration certificate
In addition for constant climate chambers HPP
► Water tank with connection hose
► Tank holder (only for appliances of size 400/750/1060, see page 24)
2.13 Optional accessories
► Converter USB to Ethernet (Fig. 8). Makes it
possible to connect the appliance's
network interface (see page 12) to the
USB port of a computer / laptop.
► Reinforced, sliding steel grids with a load
capacity of 60 kg each (for appliance size
110 and larger)
Fig. 8 Converter USB to Ethernet