D24030 | Date 11/2018 33
Operation and control
You can now
► restart the programme as described
► select another programme for processing in menu mode (see page 56) and run it as
described.
► return to manual mode. To do so, reactivate it by pressing the
activation key next to the status display, then turn the turn
control until the hand symbol
is highlighted in colour
and press the confirmation key.
5.5 Monitoring function
5.5.1 Temperature monitoring
The appliance is equipped with a multiple overtemperature protection in accordance with
DIN 12 880. This serves to avoid damage to the chamber load and/or appliance in case of a
malfunction:
► electronic temperature monitoring (TWW)
► automatic temperature monitor ( ASF)
The monitoring temperature of the electronic
temperature monitoring is measured via a separate Pt100
temperature sensor in the working chamber. Temperature
monitoring settings are made via the
ALARM display. The
settings made apply to all operating modes.
If temperature monitoring has been
triggered, this is indicated by the
temperature display: the actual tempera-
ture is highlighted in red and a warning
symbol
is shown ( Fig. 19 ). The type
of temperature monitoring triggered
(TWW in this example) is shown beneath
the temperature. If the acoustic alarm
has been activated in the menu mode
(
Sound, see page 57, indicated by the
speaker symbol in the alarm display), the
alarm is additionally signalled by an
intermittent acoustic signal, which can be turned off by pressing the confirmation key.
Information on what to do in this case are provided in the chapter Malfunctions, warning and
error messages on page 40.
Before reading how to adjust temperature monitoring (from page 35), please read the
description of the individual monitoring functions here.
Manual Mode
12.Sept.2012
13:44
ALARM
max
38.5°C
1.0K
min
35.5°C
auto
+
-
End 14:45
44h:44m
mi
00
LI
LI
FEUCHTE
30.0%rh
Set 30.0%rh
°C22.4
°CSet 37.0
TIMER
30m04h
Ende
23.Nov 13:30
°C
TEMP
TWW Set 38.5 °C
38.9
TIMER
FE
22.4
37.0
TEMP
Set
°C
°C
°C22.4
TEMP
°CSet 37.0
TIMER
°C
TEMP
TWW Set 38.5 °C
38.9
Fig. 19
Temperature monitoring triggered