32 D24030 | Date 11/2018
Operation and control
3. To start the programme, press the confir-
mation key. The programme starts. The
display shows:
the programme description (in this exam-
ple
Test 012)
the programme segment description, in
this example
Ramp 1
the current run (in case of loops)
Test 012
Ramp 1
10:4412.Sept.2012
You cannot change any parameters (e.g. the temperature) at the appliance while a
programme is running. However, the displays
ALARM and GRAPH can still be used.
Cancelling a programme
You can cancel an active programme at any
time.
1. Press the activation key to the right of
the status display. The status display is
automatically highlighted.
0
%
manueller Betrieb
%
Fr 20.10.2010 20:31
0
°
C
80
GRAPH
8
0
GRAPH
APH
Test 012
Ramp 3
10:4412.Sept.2012
2. Turn the turn control until the
stop
symbol is highlighted.
Cancel program
Test 012
10:4812.Sept.2012
3. Press the confirmation key to confirm.
The programme is cancelled.
End
Test 012
10:4912.Sept.2012
A cancelled programme cannot be resumed at the point it was cancelled. It must be
restarted from the beginning.
End of programme
The display shows End when the programme
is finished.
End
Test 012
10:4912.Sept.2012