D24030 | Date 11/2018 43
Malfunctions, warning and error messages
6.3 Power failure
In case of a power failure, the appliance operates as follows:
In manual mode
After power supply has been restored, operation is continued with the parameters set. The
time and duration of the power failure are documented in the log memory.
In timer or programme mode
In case of an interruption of the power supply of less than 60 minutes, the current pro-
gramme is continued from the point at which it was interrupted. For longer interruptions of
the power supply, all appliance functions (heating, fan etc.) are switched off.
In remote control mode:
The previous values are restored. If a programme has been initiated via remote control, it is
continued.