D24030 | Date 11/2018 49
Menu mode
7.3.6 Balance (only for model sizes 260, 400 and 750/1060)
For appliances of the sizes 260, 400 and 750/1060, application-specific correction of the heat
output distribution (balance) between the upper and lower heating groups is possible. The
adjustment range is from –50 % to +50 %.
upper heat output
upper heat output
lower heat output
-20%
lower heat output
+30%
SETUP
Balance
-20 %
2/2
SETUP
Balance
+30 %
2/2
Fig. 27 Heat output distribution (example): The –20 % (left) setting causes the lower heating
groups to emit 20 % less heat than the upper ones. The +30 % (right) setting causes the lower
heating groups to emit 30 % more heat than the upper ones. The 0 % setting restores the
default heat output distribution.
7.3.7 Remote control
In the setup entry Remote control, you can set whether the
appliance should be controlled via remote control and if
so, in which mode. These settings are available:
Off
Read Only
Write
Write+Alarm
When the appliance is in remote control mode, the
symbol appears in the temperature display. In the
settings
Write and Write+Alarm, the appliance cannot
be controlled at the ControlCOCKPIT until the remote
control has been switched off (setting Off) or set to
Read
Only
.
In order to use the remote control function, program-
ming skills and special libraries are required.
Setup
Balance
Remote Control Off
Gateway 192.168.5 .1
+30 %
2/2
23.2°C
TEMP
Set 38
.0°C