ower heat
fresh air
0
1
2
3
4
5
6
alarm set
UE/BE 200-800, SE 200-400, ULE/SLE 400-800
4
2.1 Electrical supply
50 or 60 Hz. Adequate insulated supply which incorporates earth conductor according to
EN 61010, Protection IP 20, no humidity protection according to DIN 40 050. Interference
suppression Grade N according to VDE 0875, limiting values of the classe B, part one.
A fuse 250V/15A quick is used for unit protection (in units with 400V 3N~ power supply 3
fuses).
Before current-connection please compare label on the oven and instructions of
your local current supplier.
(e.g. in Germany, DIN VDE 0100 with FI - safety device).
model volume current
consum.
power voltage ±10% weight
UE/SE 200 32 l 4,8 A 1100 W 230 V~ 28 kg
UE/SE 300 39 l 5,2 A 1200 W 230 V~ 30 kg
UE/SE/ULE/SLE 400 53 l 6,1A 1400 W 230 V~ 35 kg
UE/ULE/SLE 500 108 l 8,7 A 2000 W 230 V~ 50 kg
UE/ULE/SLE 600 256 l 10,4 A 2400 W 230 V~ 87 kg
UE/ULE/SLE 700 416 l 5,8 A 4000 W 400 V 3N~ 121 kg
UE/ULE/SLE 800 749 l 7,0 A 4800 W 400 V 3N~ 164 kg
BE 200 32 l 1,9 A 440 W 230 V~ 28 kg
BE 300 39 l 2,2 A 500 W 230 V~ 30 kg
BE 400 53 l 3,5 A 800 W 230 V~ 35 kg
BE 500 108 l 3,9 A 900 W 230 V~ 50 kg
BE 600 256 l 7,0 A 1600 W 230 V~ 87 kg
BE 700 416 l 7,8 A 1800 W 230 V~ 121 kg
BE 800 749 l 8,7 A 2000 W 230 V~ 164 kg
This equipment is intended for operation on a supply system with a system impedance
Z
max
at the transfer point (building connection) which does not exceed 0.12 Ohm. The user
has to ensure that the equipment is being operated on a supply system which meets this
requirement. If necessary, details of the system impedance can be obtained from the local
power supply authority.
Quality of material
Memmert is using stainless steel (Spec. 1.4301) for the external casing as well as
for the interior, an outstanding material because of its high stability, optimum
hygienic features and corrosion resistance against many (not all!) chemical
combinations (Attention e.g. at chlorine combinations!).
WARNING!
Before removing top cover – pull out plug !