IX 3131 II - 21
Test frequency 1 kHz
Cap.
Range
Max Display Cx Accuracy DF Accuracy Specified
Note
1 mF 0.999 mF
(*)
+/-(2.5%+5dgt)
(DF<0.1)
+/-
(5%+100/Cx+5dgt)
(DF<0.1)
after short cal.
100 µF 99.99 µF +/-(0.6%+5dgt)
(DF<0.1)
+/-
(1.2%+100/Cx+5dgt)
(DF<0.1)
after short cal.
10 µF 9.999 µF +/-(0.4%+3dgt)
(DF<0.5)
+/-
(0.4%+100/Cx+5dgt)
(DF<0.5)
-
1000 nF 999.9 nF +/-(0.4%+3dgt)
(DF<0.5)
+/-
(0.4%+100/Cx+5dgt)
(DF<0.5)
-
100 nF 99.99 nF +/-(0.4%+3dgt)
(DF<0.5)
+/-
(0.4%+100/Cx+5dgt)
(DF<0.5)
-
10 nF 9.999 nF +/-(0.4%+5dgt)
(DF<0.5)
+/-
(0.4%+100/Cx+5dgt)
(DF<0.5)
after open cal.
1000 pF 999.9 pF +/-(0.6%+5dgt)
(DF<0.1)
+/-
(1%+100/Cx+5dgt)
(DF<0.1)
after open cal.
(*) This reading can be extended to 1999 max display.
"
""
"
Notes : 1. Q value is the reciprocal of DF.
2. This specification is based on the measurement performed
at the test socket.
3. DUT & Test lead to be properly shielded to GUARD if
necessary.
4. Cx = counts of displayed C value. e. g. C = 88.88 µF than
Cx = 8888.
Inductance
Test frequency 120 Hz
Inductance
Range
Max
Display
Lx Accuracy
(DF<0.5)
DF Accuracy
(DF<0.5)
Specified
Note
10000 H 9999 H not specified not specified -
1000 H 999.9 H +/-
[0.3%+(Lx/10000)%
+5dgt]
+/-
(1%+100/Lx+5dgt)
after open cal.
100 H 99.99 H +/-
[0.3%+(Lx/10000)%
+5dgt]
+/-
(0.8%+100/Lx+5dgt)
-
10 H 9.999 H +/-
[0.3%+(Lx/10000)%
+5dgt]
+/-
(0.8%+100/Lx+5dgt)
-
1 H 999.9 mH +/-
[0.3%+(Lx/10000)%
+5dgt]
+/-
(0.8%+100/Lx+5dgt)
-
100 mH 99.99 mH +/-
[0.5%+(Lx/10000)%
+5dgt]
+/-
(1.5%+100/Lx+5dgt)
after short cal.
10 mH 9.999 mH +/-
[1.0%+(Lx/10000)%
+5dgt]
+/-
(5%+100/Lx+5dgt)
after short cal.