6.2 Dosing accuracy
Every dosing unit is subjected to a strict quality inspection prior to ship-
ment. Every dosing unit is issued a quality certificate attesting conform-
ance with the quality criteria of Metrohm.
6.2.1 Typical measurement deviation
The accuracy of dosing units can be seen in the following table. The val-
ues listed are to be regarded as typical values which can be achieved with
a 700 Dosino or an 800 Dosino.
Table 3
Typical measurement deviation of Metrohm dosing units
Cylinder volume max. systematic deviation
2 mL ± 6 µL
5 mL ± 15 µL
10 mL ± 20 µL
20 mL ± 30 µL
50 mL ± 50 µL
6.2.2 The ISO/EN/DIN standard 8655-3
The Metrohm dosing units fulfil the requirements of the ISO/EN/DIN stan-
dard 8655-3 Volume measurement instruments with pistons – Part
3: Piston burets. Metrohm guarantees that its dosing units are in com-
pliance with the following limit values at the time of shipment: