4.11 Metrohm Suppressor Module (MSM)
■■■■■■■■■■■■■■■■■■■■■■
36
■■■■■■■■
930 Compact IC Flex Oven/SeS/PP (2.930.2500)
4.11.2.3 Connecting the regeneration solution
The regeneration solution is connected to the regenerant capillary.
Connecting the regeneration solution to the peristaltic pump
For this step, you need the following accessories:
■ Accessory kit: Flex/Vario: SeS (6.5000.020)
■ Pump tubing (6.1826.420)
■ Tubing olive with filter and locking nut (6.2744.180)
■ Tubing olive (6.2744.034)
■ Tubing cartridge of the peristaltic pump
1
Prepare a tubing cartridge of the peristaltic pump for the regenera-
tion solution (see chapter 4.12.1, page 38).
2
Connect the capillary labeled regenerant to the exit of the pump tub-
ing using a pressure screw (6.2744.070).
3
Connect the PTFE capillary from the regeneration solution bottle to
the inlet of the pump tubing.