EasyManuals Logo

Metrohm OMNIS User Manual

Metrohm OMNIS
104 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #103 background imageLoading...
Page #103 background image
■■■■■■■■■■■■■■■■■■■■■■
Technical specifications
■■■■■■■■
95
Measuring resolution 0.01 µA
Potentiometric differential measure-
ment
Measuring range –2,400 mV - +2,400 mV
Measuring resolution 1.56 µV
Measuring accuracy ±1.0 mV in the measuring range
–2,000 mV - +2,000
mV
Reference conditions
Instrument status min. 30 minutes in
operation
Adjusting interval annual
Measuring accuracy applies for all measur-
ing ranges without sen-
sor error, under refer-
ence conditions, meas-
uring interval 100 ms
11.16 OMNIS Titration Module – Liquid Handling specifica-
tions
Dosing drive
Dosing resolution 100,000 steps per cylinder vol-
ume
Dosing accuracy 0.01% typically, of the dosed
volume with respect to
the cylinder volume