Command Name Data to
ILR11xx
Data from
ILR11xx
Specification
„GDB“ get energy Energy value 0.. 120 dB Indicates the amount of receiving energy.
„GNR“ get serial
number
„xxxxxxxxxx“ Serial number is sent as ASCII text
(max. 24 characters).
„GSI“ get error status „DDDDDDD“
76543212
Output of the error state,
D = “0“: no error D = “1“: error
D7: Transmitter faulty
D6: Receiver blinded or faulty
D5: Temperature warning: T < 10 °C or +70 °C
D4: Target out of range or receiver faulty
D3: Temperature error: T > +85 °C
D2: Supply voltage to low
D1: PLL unlocked
D0: always „0“
„GTE“ get
temperature
„±DDD“ DDD = Internal temperature in C°
„GVE“ get version „Fx9x $Revision X.XX$“ Software version is sent.
„GCM“ help command
get commands
All available commands All available commands are sent in text format.
„ICM“ input
continuous
mode
„0“, 1“ ACK Measurement mode setting:
„0“ = continuous measurement output,
„1“ = single measurement output
„IDO“ input offset 12000
+12000
respectively
48000..
+48000
ACK Offset setting in [mm]
or in [INCH * 100]