In the Utility menu, select "TAL" by the cursor and press
ENTER.
Use the ▼ ▲ buttons to select the desired units of dosage
and press ENTER.
If you want to esc programming and return to measurement,
press ESC for 3 seconds.
RESTORE PROGRAMMING AND CALIBRATION PARAMETERS
Attention: the Programming Reset has no influence on Language and Password.
In the Utility Menu, select "Reset" by the cursor and press
ENTER.
Use the ▼ ▲ keys to display "Yes" on Confirm. Press
ENTER.
With ▼ ▲ keys choose to restore the "Programming
Parameters" or the "Calibration Parameters". Press ENTER.
Automatically the pump restores the factory settings.
Press the ESC button for 3 seconds to exit programming.
The closing (or opening, this dipends on the programming in Utility menu, default is Normally Open
contact) of the flow input contact, free of tension, will cause:
1) the pump stops dosing
2) the green ON led blinks
3) The display shows “Flow” on the second line which alternates with the writing of the operating mode
and the flow rate.
When the flow contact returns in normal position, the pump exits from the Flow Alarm state and start
the working mode compatible with the actual inputs. If a Delay on has been set, at the end of the Flow
Alarm the pump will go on Delay on.
NOTE: The flow alarm resets the counting of the Alarm Time.
PROGRAMMING OF THE FLOW SENSOR
In the Utility Menu, select "Flow" by the cursor and press
ENTER.
Use ▼ ▲ buttons to select the type of functioning of the flow
sensor. Press ENTER.
N.O= Normally Open (default value)
N.C= Normally Close
Press the ESC button for 3 seconds to exit programming.