2 Pre-installation Instructions
ANALYZER SPACE
An unobstructed lab work space that will accommodate the following specifications is needed:
Height
Width
Depth
Weight
112 cm (44 in.)
57.25 cm (23 in.)
61 cm (24 in.)
84 kg (185 lbs)
Analyzer
Width 96.5 cm (38 in.) Approx.
Computer and Printer
1 square ft (0.1 square m) for each gas cylinder
needed for installation. For standard
installations, the cylinders must be within 6ft
(2m) of the instrument.
Gas Supply
Height
Width
Depth
17 (7 in.) Approx.
15 cm (6 in.) Approx.
31 cm (12 in.) Approx.
External Vacuum Pump
Width
Depth
50 cm (20 in.) Approx.
40 cm (16 in.) Approx.
Degas Unit (optional accessory)
Physisorption
Chemisorption
2 - 2
Flex Series Pre-installation Instructions and Checklist
350-42870-01 (Rev H) Apr 2021