GAS FOR ANALYZER TEST
To verify proper analyzer operation and to train users, Micromeritics representatives will analyze
the reference material provided in the analyzer accessories kit.
The following gases are required in order to analyze the reference materials. If these gases are
not available, Micromeritics representatives will only be able to perform a limited number of
analyzer tests during installation and operator training.
All gases require the indicated purity.
All analyzer types Chemical adsorption units
(CGA 580) N
2
99.999%
(CGA 580) He 99.999%
All gases listed to the left and:
(CGA 350) H
2
99.999%
(CGA 350) CO 99.990%
Pre-mix 10% CO in helium (required for Dynamic Chem-
ical Adsorption units and TCD units only)
Any additional gases that may be used after the installation is complete can be connected by the
Micromeritics representative.
DEGAS UNIT
Although some samples can be degassed on the analyzer, it is highly recommended that a
separate degas unit such as a Smart VacPrep is available to ensure that proper degassing
of all sample types can be performed. Using a degas unit avoids moisture and contamination in
the analyzer.
Laboratory Equipment and Supplies
Flex Series Pre-installation Instructions and Checklist
350-42870-01 (Rev H) Apr 2021
2 - 11

Other manuals for Micromeritics FLEX Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Micromeritics FLEX Series and is the answer not in the manual?

Micromeritics FLEX Series Specifications

General IconGeneral
BrandMicromeritics
ModelFLEX Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals