GAS SUPPLY CHECKLIST
Gas Supply
Initial / Date
Aclean compressed air supply at 100 ±10 psig with a 1/4 in.
(recommended) or 1/8 in. fitting on a dedicated line is required. Will
it be available at the instrument?
Y N
Are gas cylinders located within 6 ft (2 m) of where the instrument
will be installed?
Y N
Were gas regulators purchased from Micromeritics? Y N
n If No, do your gas regulators meet Micromeritics' specifications? Y N
Required Gases
Initial / Date
Are the following required gases available? The installation will not be scheduled until these
gases are available:
All instrument types:
(CGA 580) N
2
99.999%
Y N
(CGA 580) He 99.999%
Y N
Chemisorption instruments (in addition to the above gases):
(CGA 350) H
2
99.999%
Y N
(CGA 350) CO 99.990%
Y N
Pre-mix 10% CO in helium (required for Dynamic Chemisorption
units and TCD units only)
Y N
Additional Gases
Additional gases for use after installation can be connected by the Micromeritics service rep-
resentative. Please list any gases that will be available for connection during installation.
Initial / Date
Gas Supply Checklist
Flex Series Pre-installation Instructions and Checklist
350-42870-01 (Rev H) Apr 2021
3 - 3

Other manuals for Micromeritics FLEX Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Micromeritics FLEX Series and is the answer not in the manual?

Micromeritics FLEX Series Specifications

General IconGeneral
BrandMicromeritics
ModelFLEX Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals